Digital combined pH/ORP sensor Memosens CPS16D

Memosens CPS16D - Combined pH/ORP sensor with dirt-repellent PTFE diaphragm ©Endress+Hauser
Memosens CPS16D - Detail view platinum element ©Endress+Hauser
Memosens CPS16D for simultaneous pH/ORP measurement ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens glass electrode for standard applications

Memosens CPS16D is the digital all-rounder for simultaneous pH and ORP measurement offering better process control and increased quality. The sensor features a dirt-repellent PTFE diaphragm for process and environment applications with long-term stable conditions. Thanks to Memosens digital technology, CPS16D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.

  • Fördelar

    • Simultaneous measurement of pH, ORP and rH values (in rH mode) for a better process overview

    • Additional platinum electrode enables constant monitoring of reference impedance and thus sensor quality

    • Robust and resistant to poisoning thanks to large PTFE...

  • Applikationsområde

    • Long-term monitoring or limit control in processes with stable process conditions:
      - Chemical processes
      - Pulp and paper industry
      - Power stations (e.g. flue gas cleaning)
      - Incinerator plants
      - Mining
      - Wastewater

    • Water treatment:
      - Drinking water
      -...

Funktioner och specifikationer

Tillbaka

Details

  • Mätprincip

    Glass electrode

  • Application

    Standard applications in process and environment. Long-term monitoring, processes with little change.

  • Characteristic

    Very robust with ion trap and dirt repellent PTFE diaphragm.

  • Measuring principle

    Gel compact electrode with PTFE ring-diaphragm with ion trap. Pt-disc as additional ORP element.
    rH measurment and control of the reference impedance.

  • Design

    - all shaft lengths with temperature sensor, advanced gel technology - digital electrode with Memosens technology

  • Material

    Glass / PTFE

  • Dimension

    Durchmesser : 12mmSchaftlängen : 120, 225, 360 und 425mm

  • Process temperature

    max. 135°C (274°F)

  • Process pressure

    max. 16 bar (230 psi)

  • Temperature sensor

    NTC 30k

  • Connection

    Inductive, digital connection head with Memosens technology.

  • Ingres protection

    IP68

Dokumentation / manualer / programvara

Tillbehör / reservdelar

Nr
Order code
Specification