Digital combined pH/ORP sensor Memosens CPS96D

Memosens CPS96D - Digital combined pH/ORP electrode for heavily soiled media ©Endress+Hauser
Memosens CPS96D - Detail view platinum element ©Endress+Hauser
Memosens CPS96D - Digital combined pH/ORP sensor for chemical, paper and paint industries ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens glass sensor for heavily polluted media in chemical processes, paper or paint production

Memosens CPS96D is the expert for fibrous media or media with a high content of suspended solids such as dispersions, precipitation or emulsions. The open aperture guarantees the most reliable, simultaneous pH and ORP measurement for better process control and increased quality. Thanks to Memosens digital technology, CPS96D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.

  • Fördelar

    • Simultaneous measurement of pH, ORP and rH values (in rH mode) for a better process overview

    • Additional platinum electrode enables constant monitoring of reference impedance and thus sensor quality

    • Robust and resistant to poisoning thanks to stabilized...

  • Applikationsområde

    • Chemical processes

    • Pulp and paper industry

    • Contaminated media and suspended solids
      - Dispersions
      - Precipitation
      - Emulsions

    With ATEX and IEC approval for use in hazardous areas.

Funktioner och specifikationer

Tillbaka

Details

  • Mätprincip

    Glass electrode

  • Application

    - Process applications - media with high contamination: emulsions, suspensions, precipitation reactions - min. conductivity >500µS/cm

  • Characteristic

    - Rapid response - new, stabilised gel insensitive to contaminations

  • Measurement range

    pH 0-14

  • Measuring principle

    - Open aperture - double gel reference with ion trap
    Pt-disc as additional ORP element.
    rH measurment and control of the reference impedance.

  • Design

    - All shaft lengths with temperature sensor - advanced gel technology - digital electrode with Memosens technology

  • Material

    Glass

  • Dimension

    Diameter: 12 mm (0.46 inch)
    Shaft lengths: 120, 225, 360 and 425 mm
    (4.68, 8.77, 14.04 and 16,57 inch)

  • Process temperature

    max. 110°C
    (230°F)

  • Process pressure

    max. 10 bar
    (145 psi)

  • Temperature sensor

    NTC 30k

  • Connection

    Inductive, contactless connection head with Memosens technology

  • Ingres protection

    IP68

Dokumentation / manualer / programvara

Tillbehör / reservdelar

Nr
Order code
Specification