Analog disinfection sensor
CCS241

Analog disinfection sensor
CCS241 ©Endress+Hauser

Trace sensor for drinking and industrial water treatment

CCS241 is the analog sensor for trace measurement of chlorine dioxide. It delivers reliable values even if fluctuations in flow or conductivity occur and allows control of the disinfection loop so that the water is free of germs. At the same time, the sensor is designed for long calibration and service intervals reducing your maintenance effort.

  • Fördelar

    • Recalibration intervals approx. 1 to 4 months under constant operating conditions

    • Easy membrane replacement thanks to prefabricated membrane cap

    • Measured values are not affected by conductivity fluctuation

    • Measurement virtually independent of flow rate in...

  • Applikationsområde

    CCS241 measures chlorine dioxide in:

    • Drinking water
      - Trace measurement in process water and distribution networks
      - Monitoring, control and optimization of disinfection processes

    • Utilities of all industries
      - Trace measurement in all disinfection...

Funktioner och specifikationer

Tillbaka

Details

  • Mätprincip

    Chlorine dioxide

  • Application

    Drinking water, water, process

  • Characteristic

    Amperometric measurement of dissolved chlordioxide

  • Measurement range

    0,01 - 5ppm Chlordioxide

  • Measuring principle

    - closed (membrane covered) measuring cell.
    - Reduction of chlordioxide (ClO2) to chlorid at the cathode with appr. 120mV.

  • Design

    Closed amperometric 2-electrode measuring cell with PTFE membrane.

  • Material

    Sensor shaft : PVC
    Membrane : PTFE
    Membrane cap : PBT (GF30); PVDF

  • Dimension

    Diameter : appr. 25mm
    Length : 160mm

  • Process temperature

    2°C ... 45°C

  • Process pressure

    0bar - 1bar

  • Temperature sensor

    With NTC-temperature sensor

  • Connection

    Fixed cable

Dokumentation / manualer / programvara