Cerabar S PMP731

Produktbild av: Cerabar S PMP731

This product is no longer available.

Please check new generation in the section predecessor / successor or contact your local Endress+Hauser entity.

Funktioner och specifikationer

Tillbaka

Details

  • Mätprincip

    Absolute and gauge pressure

  • Characteristic / Application

    Digital pressure transmitter, piezoresistive polysilicon sensor with welded metal measuring diaphragm
    Safety by function monitoring, modular transmitter design

  • Supply / Communication

    4...20mA HART:
    11,5...45V DC
    Ex ia: 11,5...30V DC
    Ex d[ia]: 13...30V DC
    PROFIBUS PA /
    Foundation Fielbus:
    9...32V DC
    Ex ia: 9...24V DC

  • Accuracy

    0,1%

  • Long term stability

    0,1% of URL/year
    0,25% of URL/5 years

  • Ambient temperature

    -40°C...85°C

  • Process temperature

    -40°C...100°C

  • Process pressure absolute / max. overpressure limit

    600bar

  • Pressure measuring range

    1bar...400bar

  • Main wetted parts

    Alloy C
    316L
    Tantal
    PTFE-Foil

  • Process connection

    Threads (internal/flush mounted membrane)

  • Max. measurement distance

    4000m H2O

  • Communication

    4 ... 20 mA HART
    PROFIBUS PA
    FOUNDATION Fieldbus

  • Certificates / Approvals

    ATEX
    FM
    CSA
    TIIS
    SIL
    NACE
    EN10204-3.1

  • Options

    digital display
    SS- or Aluminumhousing

  • Successor

    PMP71, PMP72

  • Application limits

    Changing density

  • Mätprincip

    Absolute and gauge pressure

  • Characteristic

    Digital pressure transmitter, piezoresistive polysilicon sensor with welded metal measuring diaphragm
    Safety by function monitoring, modular transmitter design

  • Supply voltage

    4...20mA HART:
    11,5...45V DC
    Ex ia: 11,5...30V DC
    Ex d[ia]: 13...30V DC
    PROFIBUS PA /
    Foundation Fielbus:
    9...32V DC
    Ex ia: 9...24V DC

  • Reference Accuracy

    0,1%

  • Long term stability

    0,1% of URL/year
    0,25% of URL/5 years

  • Process temperature

    -40°C...100°C

  • Ambient temperature

    -40°C...85°C

  • Measuring cell

    1bar...400bar

  • Max. Turn down

    max. 100:1

  • Max. overpressure limit

    600bar

  • Process connection

    Threads (internal/flush mounted membrane)

  • Communication

    4...20mA HART
    PROFIBUS PA
    FOUNDATION Fieldbus

  • Certificates / Approvals

    ATEX
    FM
    CSA
    TIIS
    SIL
    NACE
    EN10204-3.1

  • Successor

    PMP71, PMP72

Dokumentation / manualer / programvara

Tillbehör / reservdelar

Nr
Order code
Specification
  • 14
    • Nr 14 Order code 013990-0000

      Blind plug, NPT1/2 h16.5 LD-PE

      Sealing plug NPT1/2.

    • Nr 14 Order code 015797-2000

      Blind plug, M20 h12/6, LD-PE

      Sealing plug M20.

    • Nr 14 Order code 016137-0012

      Blind plug, G1/2A h12/6 HD-PE

      Sealing plug G1/2.

  • 21
    • Nr 21 Order code 013982-0000

      O-ring 61,6 x 2,62 NBR 70 +PTFE

      Usage: cover T4.

  • 57
    • Nr 57 Order code 52009800

      Flat gasket 18,5x23,9x1,5 FKM, 5 pieces

      Process gasket FKM Viton. Usage: Cerabar M/S, Cerabar PMP11/PMP21, metal cell.

  • 70
    • Nr 70 Order code 919806-0000

      Mount.set, wall-/pipe-2", VA, lipht sep

      Usage: wall/pipe mounting. Material: 304.

  • Ersättare