Endress+Hauser-koncernen

Mediacenter

Event

En karriär hos Endress+Hauser

Support

Online Shop

Ange dina personliga inloggningsuppgifter för att komma till online-shoppen och lägg din order när som helst, dygnet runt. Här hittar du också prisuppgifter och leveranstider och kan skicka offertförfrågningar.

Online Shop

E-direct Shop

E-direct - Beställ förkonfigurerade fältinstrument och systemkomponenter snabbt och enkelt online.

E-direct Shop

Vibronic
Point level switch
Liquiphant FTL71

Point level switch with extension tube for liquids in high-temperature applications

  • Liquiphant FTL71 ©Endress+Hauser
  • Vibronic measuring principle

  • Liquiphant with permanent function check is the safe choice for point level detection in liquids

Liquiphant FT71 is a point level switch with extension tube for use in hazardous areas with all international certificates. FTL71 is especially designed for applications with high process temperatures up to 280°C and can be used up to SIL2, in homogeneous redundancy up to SIL3. An integrated second line of defense offers the highest degree of safety. Reliable measurement values, not affected by: changing media properties, flow, turbulences, gas bubbles, foam, vibrations or build-up.

Fördelar

  • Use in safety systems requiring functional safety to SIL2 in accordance with
    IEC 61508/IEC 61511-1

  • With welded gas-tight feed-through maximum safety in the event of damaged sensor

  • Wide variety of electronics, e.g. NAMUR, relay, thyristor, PFM signal output: the right connection for every process control system

  • Large number of process connections to choose from: universal usage

  • No calibration: quick, low-cost start-up

  • No mechanically moving parts: no maintenance, no wear, long operating life.
    Monitoring of fork for damage: guaranteed function

  • FDA approved materials (PFA Edlon)

Applikationsområde

Point level switch for all liquids.

  • Process connections: Threads and flanges

  • Temperature: -60 to +280°C (-76 to +540°F)

  • Pressure: -1 to +100bar (-14.5 to +1,450psi)

  • Sensor material: 316L, Alloy (PFA and Enamel on request)

  • International explosion protection certificates, overfill prevention WHG, SIL, marine approvals

Funktioner och specifikationer

Dokumentation/manualer/programvara

Tillbehör/reservdelar

Föregångare / Ersättare

Föregångare: Rod probe 11500Z, Rod probe 11500ZM, CS5603, CS5613

Kontakt

Endress+Hauser AB
Box 1486 171 28 Solna , Sverige
Tel.: +46 (0)8 555 116 00
Fax: +46 (0)8 555 116 55
Skicka e-post www.se.endress.com

Relaterade nyheter

Online-verktyg

Den här webbplatsen använder cookies. Genom att fortsätta använda webbplatsen godkänner du att vi använder cookies. Läs mer här Stäng