Vibronic
Point level detection
Liquiphant FTL70

Point level switch for liquids in
high-temperature applications

  • Liquiphant FTL70 ©Endress+Hauser
  • Liquiphant FTL70 application ©Endress+Hauser
  • Vibronic measuring principle

  • Liquiphant with permanent function check is the safe choice for point level detection in liquids

Liquiphant FT70 is a point level switch for use in hazardous areas with all international certificates. FTL70 is especially designed for applications with high process temperatures up to 280°C and can be used for Safety Instrumented Systems up to SIL2, in homogeneous redundancy up to SIL3. An integrated second line of defense offers the highest degree of safety. Reliable measurement values, not affected by: changing media properties, flow, turbulences, gas bubbles, foam, vibrations or build-up.

Fördelar

  • Use in safety systems requiring functional safety to SIL2 in accordance with
    IEC 61508/IEC 61511-1

  • With welded gas tight feedthrough maximum safety in the event of damaged sensor

  • Wide variety of electronics, e.g. NAMUR, relay, thyristor, PFM signal output: the right connection for every process control system

  • Large number of process connections to choose from: universal usage

  • No calibration: quick, low-cost start-up

  • No mechanically moving parts: no maintenance, no wear, long operating life.
    Monitoring of fork for damage: guaranteed function

  • FDA approved materials (PFA Edlon)

Applikationsområde

Point level switch for all liquids.

  • Process connections: Threads and flanges

  • Temperature: -60 to +280°C (-76 to +540°F)

  • Pressure: -1 to +100bar (-14.5 to +1,450psi)

  • Sensor material: 316L, Alloy (PFA and Enamel on request)

  • International explosion protection certificates, overfill prevention WHG, marine approvals, SIL

Funktioner och specifikationer

Dokumentation/manualer/programvara

Tillbehör/reservdelar

Föregångare / Ersättare

Föregångare: Rod probe 11500Z, Rod probe 11500ZM, CS5603, CS5613